Related Experiment Video
Updated: May 10, 2026

Magnetic Resonance Imaging of Multiple Sclerosis at 7.0 Tesla
Published on: February 19, 2021
Effect of insulating layer material on RF-induced heating for external fixation system in 1.5 T MRI system
Yan Liu1, Wolfgang Kainz, Songsong Qian
1Department of Electrical and Computer Engineering, University of Houston, Houston , TX , USA .
Abstract:
The radio frequency (RF)-induced heating is a major concern when patients with medical devices are placed inside a magnetic resonance imaging (MRI) system. In this article, numerical studies are applied to investigate the potentials of using insulated materials to reduce the RF heating for external fixation devices. It is found that by changing the dielectric constant of the insulation material, the RF-induced heating at the tips of devices can be altered. This study indicates a potential technique of developing external fixation device with low MRI RF heating.