Differences in blood pressure levels obtained by auscultatory and oscillometric methods
M G Weaver1, M K Park, D H Lee
1Department of Pediatrics, University of Texas Health Science Center, San Antonio 78284.
American Journal of Diseases of Children (1960)
|August 1, 1990
Summary
Blood pressure readings from the Dinamap Monitor differ significantly from the traditional auscultatory method in children. This study highlights the need for updated normative data for oscillometric blood pressure measurements.
Area of Science:
- Pediatric Health
- Biomedical Engineering
- Clinical Measurement
Background:
- The conventional auscultatory method is standard for blood pressure measurement.
- Oscillometric devices like the Dinamap Monitor offer an automated alternative.
- Accurate blood pressure monitoring in children is crucial for identifying hypertension.
Purpose of the Study:
- To compare blood pressure measurements obtained by the auscultatory method and the Dinamap Monitor in children.
- To assess the suitability of current normative auscultatory blood pressure levels for oscillometric readings.
- To provide a method for converting Dinamap readings to estimated auscultatory values.
Main Methods:
- Compared auscultatory and Dinamap Monitor blood pressure measurements in 381 children (ages 10-13).
- Used triplicate measurements with a cuff bladder width of 40%-50% of arm circumference.
- Calculated mean differences and developed predictive equations for systolic and diastolic pressures.
Main Results:
- The auscultatory method yielded systolic pressures 6.4 mm Hg lower and diastolic pressures 8.7 mm Hg higher than the Dinamap Monitor.
- Significant discrepancies suggest current auscultatory norms are inappropriate for Dinamap Monitor data.
- Predictive equations were derived to estimate auscultatory blood pressure from Dinamap readings.
Conclusions:
- Published auscultatory blood pressure norms are not directly applicable to measurements from the Dinamap Monitor in children.
- The study provides equations to bridge the gap between oscillometric and auscultatory measurements.
- New normative data specific to oscillometric devices are recommended for pediatric populations.