Purpose:
This study evaluated the effect of fatigue loading on the screw joint stability of a zirconium abutment connected to an external hexagon implant in vitro.
Materials And Methods:
Fifteen titanium and 15 zirconia abutments of 3 different heights (5, 8, and 11 mm) were connected to external titanium implants with titanium screws. A torque gauge was used to measure the reverse torque values before and after loading. An air cylindrical loading device was used to simulate mastication at a 45-degree angle to the longitudinal axis of the implant.
Results:
There were significant differences (P<0.05) before and after the loading of titanium (5mm) and zirconia (5, 8, and 11 mm) abutments.
Conclusion:
Zirconia abutments for external hexagon implants had durability rates similar to those of titanium abutments after repeating load on the reverse torque of the abutment screw, indicating that the zirconia abutment could be reliably used instead of the titanium abutment.
Related Concept Videos
Fatigue
Frictional Forces on Screws
A jack with a square-threaded screw is a mechanical device used to lift heavy loads by applying a force at its handle. When the force is applied, the screw turns, raising the load. The screw can...
Fatigue Strength of Concrete
Stresses under Combined Loadings
The process begins by slicing the tube at critical points and analyzing the internal forces and stress components at these sections, focusing on the centroid. Normal stresses, generated by axial forces and bending moments, are either compressive or tensile and vary across the section from...
General Case of Eccentric Axial Loading
Consider a member subjected to equal and opposite forces that are applied along a line that does not coincide with the member's neutral axis. In unsymmetrical bending,...
Normal Strain under Axial Loading