Neonatal blood pressure monitoring: visual assessment is an unreliable method for selecting cuff sizes
A Devinck1, H Keukelier, I De Savoye
1Faculty of Medicine and Health Sciences, Ghent University, Ghent, Belgium.
Aim:
To compare current practice of cuff size selection for noninvasive blood pressure measurement in a single-centre, tertiary-level neonatal intensive care unit (visual assessment of bladder width/limb length closest to 2/3) with common recommendations for appropriate cuff selection.
Methods:
Visual assessment of the appropriate cuff size ('2/3 rule') for upper arm, forearm and calf in 103 neonates (309 cuff selections) was compared with the following recommendations: (i) Method A - guidelines of the cuff manufacturer, (ii) Method B - cuff width/limb circumference ratio 0.44-0.60 and (iii) Method C - cuff width/limb length ratio closest to 0.66.
Results:
The upper arm cuff size was correctly chosen in 84% of cases (Method A), 43% (Method B) and 56% (Method C). The forearm cuff size was correctly chosen in 94% of cases (Method A), 68% (Method B) and 54% (Method C). The calf cuff size was correctly chosen in 96% of cases (Method A), 72% (Method B) and 63% (Method C).
Conclusion:
The accuracy of selecting cuff size by visual assessment is low. Further research on accurate cuff selection for neonates, including at the forearm and calf, is warranted.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Pre-Procedural Guidelines for Assessing Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessing Blood pressure in the Leg
Preparation:
Assessment of blood pressure in brachial artery(two-step method)