Effects of misalignment on static torsional strength of anterior cervical plate systems
Kingsley R Chin1, Vanessa B R Cumming, MeLeah Henson
1Institute for Modern and Innovative Surgery, 1100 W. Oakland Park Blvd, Suite #3, Fort Lauderdale, FL 33311, USA.
Background Context:
There is little understanding of cervical plate misalignment as a risk factor for plate failure at the plate-screw-bone interface.
Purpose:
To assess the torsional strength and mode of failure of cervical plates misaligned relative to the midsagittal vertical axis.
Study Design:
Plastic and foam model spine segments were tested using static compression and torsion to assess effects of misaligned and various lengths anterior cervical plate (ACPs).
Methods:
Different length ACPs and cancellous fixed angle screws underwent axial torsional testing on a servo-hydraulic test frame at a rate of 0.5°/s. A construct consisted of one ACP, four screws, one ultrahigh-molecular weight polyethylene inferior block, and one polyurethane foam superior block. Group 1 had ACPs aligned in the midsagittal vertical axis, group 2 plates were positioned 20° offset from the midline, and group 3 had the ACP shifted 5 mm away and 20° offset from midline. Torques versus angle data were recorded. The failure criterion was the first sign of pullout determined visually and graphically.
Results:
Group 1 had a more direct screw pullout during failure. For the misaligned plates, failure was a combination of the screws elongating the holes and shear forces acting between the plate and block. The misaligned plates needed more torque to failure. The failure torque was 50% reduced for the longer versus the shorter plates in the neutral position. Graphically shown initial screw slippage inside the block preceded visual identification of slippage in some cases.
Conclusions:
We observed different failure mechanisms for neutral versus misaligned plates. Clinically, misalignment may have the benefit of needing more torque to fail. Misalignment was a risk factor for failure of the screw-bone interface, especially in longer plate constructs. These comparisons of angulations may be a solid platform for expansion toward a more applicable in vivo model.
Related Concept Videos
Eccentric Axial Loading in a Plane of Symmetry
Torsion of Noncircular Members
Plastic Deformations of Members with a Single Plane of Symmetry
Unsymmetric Bending - Angle of Neutral Axis
When a bending moment is applied at an angle θ concerning the vertical axis of a symmetrical member, it can be resolved into components along the member's principal centroidal axes. The...
Stability of structures
General Case of Eccentric Axial Loading
Consider a member subjected to equal and opposite forces that are applied along a line that does not coincide with the member's neutral axis. In unsymmetrical bending,...