Related Experiment Video
Updated: May 10, 2026

Infant Auditory Processing and Event-related Brain Oscillations
Published on: July 1, 2015
Detectability of newborn chirp-evoked ABR in the frequency domain at different stimulus rates
Mario Cebulla1, Ekkehard Stürzebecher
1* Comprehensive Hearing Center (CHC), Department of Otorhinolaryngology, Plastic, Aesthetic and Reconstructive Head and Neck Surgery, Julius Maximilian-University Hospitals , Würzburg , Germany.
Objective:
The present paper reports the results of a study on a group of newborns whose chirp-evoked auditory brainstem responses (ABR) were recorded at different repetition rates. The study was aimed at finding an optimum repetition rate for the chirp stimulus that facilitates a short response detection time in the frequency domain.
Design:
Chirp-evoked ABR were recorded at six different stimulation rates between 20/s and 100/s. The stimulation level was 35 dB nHL. The mean harmonic related SNR was calculated, and the response detection time and rate were assessed using a detection algorithm.
Study Sample:
A group of 80 sleeping and 27 awake newborns with normal hearing were included.
Results:
The highest mean harmonic-related SNR in both groups was found at 60/s. The signal-to-noise ratio (SNR) was significantly smaller for the awake newborns. A significantly shorter detection time can be achieved for both newborn groups at 60/s compared to the other rates tested. Response detection in the group of awake newborns profited more from the change to 60/s compared to the sleeping newborns.
Conclusions:
It can be expected that with this optimized stimulus paradigm a shorter test time and a higher specificity of the hearing screening can be reached.