Mechanical and electrical equipment interference provokes a misleading Neurally Adjusted Ventilatory Assist (NAVA)
Y Somers1, W Verbrugghe, P G Jorens
1Department of Critical Care Medicine, Antwerp University Hospital, University of Antwerp, Antwerp, Belgium - Philippe.Jorens@uza.be.
Background:
Neurally Adjusted Ventilatory Assist (NAVA) offers synchronized proportional pressure in accordance with the electrical activity of the diaphragm (EAdi). NAVA relies on the EAdi to trigger the respiratory cycle and then adjusts the ventilatory assist to the neural drive. The technique necessitates a catheter with bipolar microelectrodes positioned near the crural diaphragm where this signal can be captured. Capturing a reliable EAdi signal is a condition sine qua non for using NAVA as a mode of ventilation. The displayed signal represents the sum of the electrical activity of the muscle action potential of the diaphragm and is expressed in microvolts.
Methods:
A technical note illustrated by a case series in the intensive care unit (ICU) of a tertiary referral hospital with experience using NAVA.
Results:
Here, we report on three separate cases in which the use of an intra-aortic balloon catheter, a pacemaker and a heating device all resulted in a distortion of the EAdi signal, despite good positioning of the catheter. In a fourth case, we observed internal interference from leaking cardiac electrical activity due to the malpositioning of the EAdi catheter in a patient with atrial fibrillation.
Conclusion:
We illustrate that the detection and therefore interpretation of the EAdi signal during NAVA can be influenced by mechanical and electrical interference by other equipment used in the ICU or from endogenous leaking cardiac activity.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Neural Control of Respiration
Respiratory Centers in the Brainstem
Two primary areas comprise the respiratory center: the medullary respiratory center in the medulla oblongata and the pontine respiratory group in the pons. The...
Errors occurring during blood pressure monitoring
Several factors...