Related Experiment Video
Updated: May 10, 2026

Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
Use of continuous infusion pumps during radiation treatment
Kate Bak1, Eric Gutierrez, Elizabeth Lockhart
1Cancer Care Ontario, Toronto, Canada. Kate.Bak@cancercare.on.ca
Introduction:
Despite increasing chemoradiotherapy treatment, there is a paucity of information regarding the effects of radiation exposure on ambulatory infusion pumps used to deliver chemotherapy or other essential medications. The aim of this overview is to present the available evidence on this subject, heighten awareness within the clinical community, provide considerations for minimizing possible negative effects on patient care, and encourage the monitoring of infusion devices after exposure to radiation or electromagnetic interference.
Methods:
Published literature was systematically searched using MEDLINE and EMBASE; gray literature was searched using Google and an environmental scan of relevant Web sites. A multidisciplinary working group reviewed the compiled evidence, and a draft of the document was sent to health professionals from various disciplines for an external review.
Results:
Four reports and three manufacturer device alerts were identified that suggest a risk of pump malfunction as a result of radiation exposure. The estimated cumulative dose at which pump failure has been reported ranges from 28.5 to 42 Gy; however, additional clinical investigations should be undertaken. Pump relocation, pump shielding, and assessment of the pump after radiation exposure are most commonly suggested to minimize pump malfunction related to radiation exposure. A list of additional considerations is offered for those developing institution specific policies and procedures based on the available evidence and expert consensus.
Conclusion:
The varied and unpredictable results of radiation exposure on infusion devices suggest that additional testing should be carried out to determine the limits of dose exposure and to raise awareness around this patient safety issue.
Related Concept Videos
Drug Accumulation During Multiple Dosing: Intermittent IV Infusions
Continuous Renal Replacement Therapy
Extracorporeal Removal of Drugs: Continuous Renal Replacement Therapy
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations
Oral Drug Delivery Systems: Continuous-Release Systems