Related Experiment Video
Updated: May 10, 2026

Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
Video-assisted pericardioscopic surgery for epimyocardial lead implantation
Nima Hatam1, Jan Spillner, Eva Maria Nosthoff
1Department of Cardiovascular and Thoracic Surgery, University Hospital, Medical Faculty RWTH Aachen, Aachen, Germany. nhatam@ukaachen.de
Purpose:
Video-assisted pericardioscopic surgery (VAPS) for epimyocardial lead implantation has demonstrated positive acute results concerning the safety and degree of freedom inside the pericardium. We evaluated the employment of a newly developed trocar for pericardioscopy with regard to long-term effects and feasibility of reoperation.
Description:
Eight adult sheep were divided into three groups. In two animals, VAPS was used exclusively. All other animals received four small-caliber epicardial leads through VAPS. After 6 and 12 months (n = 3 each), reoperation was conducted for reevaluation of entry site, intrapericardial adhesions, lead position, and morphology of the implantation site.
Evaluation:
Reentry close to the previous entry site proves unproblematic. Adhesions were mild to moderate in the immediate area of the implanted leads. Throughout the follow-up, pacing parameters were satisfactory. Lead dislodgement occurred in 1 of 24 leads. The deployment of small-caliber flexible endoscopes through the new trocar provided sufficient navigation, stability, and maneuverability.
Conclusions:
Reoperation from the same subxiphoid approach proved feasible. Lead removal and reimplantation were feasible at both 6 months and 12 months after initial implantation. The intrapericardial adhesions caused by VAPS alone are mild.
More Related Videos
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
Related Concept Videos
Cardiomyopathy V: Interprofessional Care
Cardiac Catheterization III: Left Heart Catheterization
Cardiac Catheterization II: Right Heart Catheterization