Patient specific pre-treatment QA verification using an EPID approach
Luis A Vazquez Quino1, Xiaoming Chen, Mathew Fitzpatrick
1Cancer Therapy and Research Center, Department of Radiation Oncology, 7979 Wurzbach Rd., University of Texas Health Science Center at San Antonio, TX, 78229, USA. papanikolaou@uthscsa.edu.
A new software, MU-EPID, verifies patient-specific intensity modulated radiation therapy (IMRT) quality assurance using electronic portal imaging device (EPID) measurements. This tool accurately reconstructs fluence maps for reliable pre-treatment verification.
Area of Science:
- Medical Physics
- Radiation Oncology
- Software Engineering
Background:
- Intensity modulated radiation therapy (IMRT) requires rigorous patient-specific quality assurance (QA) for safe and effective treatment.
- Electronic portal imaging devices (EPIDs) offer a method for measuring radiation fluence during treatment delivery.
- Accurate conversion of EPID measurements to treatment planning system (TPS) equivalent fluence maps is crucial for QA.
Purpose of the Study:
- To validate the MU-EPID software for patient-specific pre-treatment QA of IMRT.
- To assess the accuracy of fluence map reconstruction from EPID measurements using the MU-EPID software.
- To evaluate the software's capability in providing comprehensive 3D and 2D dosimetric analysis.
Main Methods:
- Developed MU-EPID software to convert EPID-acquired DICOM images to TPS-equivalent fluence maps.
- Processed Varian, Elekta, and Siemens EPID DICOM images.
- Validated the software using 20 IMRT plans from various treatment sites with a Varian a-Si 1000 EPID and Pinnacle TPS.
- Performed comparative analyses of isodose distributions, dose profiles, DVHs, and gamma index.
Main Results:
- The MU-EPID software successfully converted EPID images to equivalent fluence maps.
- Gamma analysis showed an average passing rate of 97.44% using a 3% and 3 mm criterion.
- Comparisons of isodose, DVH, and dose profiles demonstrated good agreement between calculated and reconstructed plans.
- The software supports Varian, Elekta, and Siemens EPID DICOM image formats.
Conclusions:
- The MU-EPID software is a clinically viable tool for patient-specific IMRT QA.
- It provides accurate 3D dosimetric evaluation through DVH comparison.
- Offers a 2D gamma analysis option for comprehensive treatment verification.