Neonatal self-inflating bags: achieving titrated oxygen delivery using low flows: an experimental study
Arun Sasi1, Natarajan Chandrakumar, Ashok Deorari
1Division of Neonatology, Department of Pediatrics, All India Institute of Medical Sciences, New Delhi, India.
Neonatal resuscitation bags (SIBs) can deliver nearly 40% oxygen at low flow rates and clinically used pressures. Increasing oxygen flow and adding a reservoir can increase delivery up to 99%.
Area of Science:
- Neonatal Resuscitation
- Respiratory Physiology
- Medical Device Engineering
Background:
- Self-inflating resuscitation bags (SIBs) are crucial for neonatal resuscitation.
- Optimizing oxygen delivery (dFiO2) from SIBs, especially at low flow rates, is essential for clinical practice.
- Understanding the impact of various parameters on dFiO2 is critical for improving patient outcomes.
Purpose of the Study:
- To determine the delivered oxygen concentration (dFiO2) during manual inflations using neonatal SIBs.
- To investigate dFiO2 at oxygen flow rates below 1 L/min.
- To assess the influence of different flow rates, peak inspiratory pressures (PIP), and inflation rates on dFiO2.
Main Methods:
- Experimental study involving 216 sets of manual inflations with two types of neonatal SIBs.
- Tested various oxygen flow rates (0.2-5.0 L/min), PIP (10-25 cmH2O), and inflation rates (30-60/min).
- Evaluated dFiO2 with and without an oxygen reservoir using a leak-proof circuit, pressure transducer, O2 analyzer, and test lung.
Main Results:
- Mean dFiO2 ranged from 75% to 93% at 5 L/min without reservoirs.
- At 1 L/min, SIBs delivered 65-85% oxygen.
- dFiO2 decreased to approximately 40% at 0.2 L/min with PIP of 20-25 cmH2O and inflations of 40-60/min.
Conclusions:
- Neonatal SIBs can achieve nearly 40% dFiO2 at clinically relevant pressures and rates using low oxygen flows.
- A graded increase in dFiO2 from 40% to 99% was observed with increasing flow rates (0.2-5 L/min) and the addition of a reservoir.
- Even at low flows, achieving dFiO2 below 40% was not possible with the tested configurations.
Related Concept Videos
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils, connected...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes: