Related Experiment Video For cone beam computed tomography
Updated: May 10, 2026

Reliability of Artificial Intelligence-Based Cone Beam Computed Tomography Integration with Digital Dental Images
Published on: February 23, 2024
Comparison between cone beam computed tomography and multislice computed tomography in diagnostic accuracy of
Amir Eskandarlou1, Jalal Poorolajal, Ahmad Reza Talaeipour
1Department of Dentomaxillofacial Radiology, School of Dentistry, Hamadan University of Medical Sciences, Hamadan, Iran.
Background:
The aim of this study was to compare the accuracy of cone beam computed tomography (CBCT) and multislice computed tomography (MSCT) in diagnosing factitious fractures in maxillofacial region as well as to investigate the influence of field of view (FOV) size on the interpretation of maxillofacial fractures in CBCT.
Methods:
Simulated fractures were created in the right and left sides of five dried human skulls using a micro-saw 20 mm blade without displacement. The skulls were scanned with a spiral 16-slice MSCT scanner and a CBCT device with two FOVs (6 × 6, 15 × 15). Three observers assessed and interpreted the images judging whether fracture was present or absent. The sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), and the area under the receiver operating characteristic (ROC) curve were calculated.
Results:
The sensitivity of CBCT 6 × 6, CBCT 15 × 15, and MSCT was 87.5%, 70.7%, and 49.7% and the specificity of three modalities was 91.7%, 88.0%, and 98.3%, respectively. The area under ROC curve was 0.90 (95% CI: 0.85, 0.94) for CBCT with FOV 6 × 6, 0.79 (95% CI: 0.75, 0.85) for CBCT with FOV 15 × 15, and 0.74 (95% CI: 0.69, 0.79) for MSCT. The Kappa value for interobserver agreement was highest for CBCT 6 × 6 among imaging modalities.
Conclusions:
CBCT with small FOV had higher diagnostic accuracy in detecting simulated maxillofacial fractures than MSCT. However, additional evidence based on in vivo studies is needed to confirm the superiority of CBCT on MSCT in diagnosis of maxillofacial traumatic fractures.
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...