An Anesthesia, Surgery, and Harvest Method for the Evaluation of Transpedicular Screws Using an In Vivo Porcine Lumbar Spine Model
Published on: May 31, 2017
Biodegradable device applied in flatfoot surgery: comparative studies between clinical and technological aspects of
Barbara Ruozi1, Daniela Belletti, Giuseppe Manfredini
1Department of Life Sciences, University of Modena and Reggio Emilia, Via Campi 183, 41100 Modena, MO, Italy. barbara.ruozi@unimore.it
Abstract:
Poly-L-lactide (PLLA) is one of the most used polymers for biomedical application; its use in sutures and other implants has been widely investigated. Although the knowledge of PLLA biodegradation and biocompatibility features is deep, PLLA screws used to correct the flat foot deformity have deserved attention since they are not degraded in most of cases after a long period of years (3-7) from the implantation. In this article, a clinical and radiological evaluation (NMR, histological and clinical outcomes) on patients was correlated with physico-chemical characterization (by SEM, DSC, GPC and XRD analysis at different temperatures) on both native and patient-recovered screws together with the theoretical degradation processes of PLLA-based implants. The data demonstrated the need for crossing the biodegradation and bioabsorption of the polymer with the characteristics of both the device (geometry, structure and fabrication process) and the implantation site.