Comparison of freehand-navigated and aiming device-navigated targeting of liver lesions
Background:
Accurate needle placement is crucial for the success of percutaneous radiological needle interventions. We compared three guiding methods using an optical-based navigation system: freehand, using a stereotactic aiming device and active depth control, and using a stereotactic aiming device and passive depth control.
Methods:
For each method, 25 punctures were performed on a non-rigid phantom. Five 1 mm metal screws were used as targets. Time requirements were recorded, and target positioning errors (TPE) were measured on control scans as the distance between needle tip and target.
Results:
Time requirements were reduced using the aiming device and passive depth control. The Euclidian TPE was similar for each method (4.6 ± 1.2-4.9 ± 1.7 mm). However, the lateral component was significantly lower when an aiming device was used (2.3 ± 1.3-2.8 ± 1.6 mm with an aiming device vs 4.2 ± 2.0 mm without).
Discussion:
Using an aiming device may increase the lateral accuracy of navigated needle insertion.