Successful retrieval of a fractured and embolised Judkin's catheter during a coronary angiogram
Soumya Patra1, K S Sadananda, Navin Agrawal
1Department of Cardiology, Sri Jayadeva Institute of Cardiovascular Sciences & Research, Bangalore, Karnataka, India. dr_soumyapatra@rediffmail.com
Insights
A fractured Judkin's right catheter segment was successfully retrieved from a patient's ascending aorta using a snare catheter. This retrograde approach through the left femoral artery provided a safe and effective solution for removing the complex arterial foreign body.
Area of Science:
- Cardiology
- Interventional Cardiology
- Medical Device Retrieval
Background:
- A 69-year-old female patient presented with an anterior wall myocardial infarction.
- Coronary angiography indicated severe stenosis in the left main coronary artery, left anterior descending artery, and left circumflex artery.
Observation:
- During a cardiac procedure, a Judkin's right catheter fractured approximately 15 cm from its distal end.
- The broken catheter segment was located in the ascending aorta, with the proximal end in the left subclavian artery.
Findings:
- An initial attempt to retrieve the distal catheter segment using a snare catheter via the right common iliac artery was unsuccessful due to the fragment's curvature.
- A retrograde approach utilizing a snare catheter through an arterial sheath in the left femoral artery successfully removed the fractured catheter segment.
Implications:
- This case highlights the successful application of a retrograde snare catheter technique for retrieving a complex, large fractured catheter segment from the aorta.
- The described method offers a viable alternative for managing similar intra-arterial foreign body complications during interventional procedures.
Abstract:
We report the successful retrieval of a large broken segment of right Judkin's catheter using a snare catheter by a retrograde approach. A 69-year-old woman presented with an anterior wall myocardial infarction. A coronary angiogram revealed severe stenosis of distal left main coronary artery, ostio-proximal portion of the left anterior descending artery and the ostial left circumflex artery. During right shoot, the Judkin's right catheter was broken at almost 15 cm from the distal end. The distal end of the broken catheter was lying in the ascending aorta, whereas proximal end was in the left subclavian artery. First, we snared the distal end and pulled out the catheter in right common iliac artery but owing to the curvature of the fragment, this attempt was unsuccessful. Finally, we were successful in removing the fractured segment retrogradely through the arterial sheath placed in left femoral artery.