Patient injuries from anesthesia gas delivery equipment: a closed claims update
Sonya P Mehta1, James B Eisenkraft, Karen L Posner
1* Fellow, ‡ Research Professor, Department of Anesthesiology and Pain Medicine, University of Washington, Seattle, Washington. † Professor, Department of Anesthesiology, Icahn School of Medicine at Mount Sinai, New York, New York. § Professor, Departments of Anesthesiology and Pain Medicine, and Neurological Surgery (Adjunct), University of Washington.
Background:
Improvements in anesthesia gas delivery equipment and provider training may increase patient safety. The authors analyzed patient injuries related to gas delivery equipment claims from the American Society of Anesthesiologists Closed Claims Project database over the decades from 1970s to the 2000s.
Methods:
After the Institutional Review Board approval, the authors reviewed the Closed Claims Project database of 9,806 total claims. Inclusion criteria were general anesthesia for surgical or obstetric anesthesia care (n = 6,022). Anesthesia gas delivery equipment was defined as any device used to convey gas to or from (but not involving) the airway management device. Claims related to anesthesia gas delivery equipment were compared between time periods by chi-square test, Fisher exact test, and Mann-Whitney U test.
Results:
Anesthesia gas delivery claims decreased over the decades (P < 0.001) to 1% of claims in the 2000s. Outcomes in claims from 1990 to 2011 (n = 40) were less severe, with a greater proportion of awareness (n = 9, 23%; P = 0.003) and pneumothorax (n = 7, 18%; P = 0.047). Severe injuries (death/permanent brain damage) occurred in supplemental oxygen supply events outside the operating room, breathing circuit events, or ventilator mishaps. The majority (85%) of claims involved provider error with (n = 7) or without (n = 27) equipment failure. Thirty-five percent of claims were judged as preventable by preanesthesia machine check.
Conclusions:
Gas delivery equipment claims in the Closed Claims Project database decreased in 1990-2011 compared with earlier decades. Provider error contributed to severe injury, especially with inadequate alarms, improvised oxygen delivery systems, and misdiagnosis or treatment of breathing circuit events.
Related Concept Videos
Inhalational Anesthetics: Overview
Cardiopulmonary Resuscitation II: ACLS Airway Management
General Anesthesia: Overview
General anesthesia induces unconsciousness in the whole body, while the others target specific areas or sensations. It is administered to minimize adverse effects, maintain...
Parenteral Anesthetics: Overview
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Endoscopic Procedures IV: Sigmoidoscopy and Laproscopy
Sigmoidoscopy
Sigmoidoscopy is a diagnostic procedure that uses a flexible sigmoidoscope equipped with a light source and camera to examine the rectum and sigmoid colon. The procedure involves inserting the tube through the anus...