Related Experiment Video
Updated: May 9, 2026

Optimizing Magnetic Force Microscopy Resolution and Sensitivity to Visualize Nanoscale Magnetic Domains
Published on: July 20, 2022
Imaging near metal: The impact of extreme static local field gradients on frequency encoding processes
Kevin M Koch1, Kevin F King, Michael Carl
1GE Healthcare, Magnetic Resonance Division, Waukesha, Wisconsin, USA.
Purpose:
Magnetic resonance imaging capabilities in the direct vicinity of metallic devices have substantially improved with the recent development of three-dimensional multispectral imaging (3D-MSI) methods. When imaging near metallic hardware, the bulk image distortions in 3D-MSI techniques are reduced to the single-pixel level. However, commonly utilized MSI techniques are ultimately limited by frequency-encoding processes and reveal a class of residual intensity-based susceptibility artifacts that have yet to be formally analyzed.
Methods:
Empirical measurements and simulation techniques are utilized to study the static local magnetic field gradients induced by metal implants and their general impact on frequency-encoding processes. The specific consequences of these gradients on 3D-MSI approaches are also analyzed using empirical and simulated approaches.
Results:
Close agreements between empirical and simulated measurements clearly demonstrate the effects of strong local gradients on frequency-encoded imaging capabilities near metallic implants.
Conclusions:
3D-MSI techniques can enable substantially enhanced magnetic resonance imaging capabilities near metallic implants. However, strong local static field gradients generate residual artifacts whose direct mitigation are ultimately limited by frequency encoding processes. Applications of 3D encoding strategies or additional post processing may be required to further reduce residual artifacts in multispectral images near metal implants.
Related Concept Videos
Atomic Nuclei: Nuclear Relaxation Processes
Induced Electric Fields: Applications
Magnetic Field due to Moving Charges
Consider a point charge moving with a constant velocity. Like the electric field, the magnetic field at any point is directly proportional to the magnitude of the charge and inversely proportional to the square of the distance between the source point and the field point. However, unlike the electric field, the magnetic field is always perpendicular to the plane containing the line...
Magnetic Field Due To A Thin Straight Wire
Magnetic Resonance Imaging
Biasing of Metal-Semiconductor Junctions
In Schottky junctions, where the semiconductor is n-type, applying a positive voltage to the metal relative to the semiconductor reduces its Fermi...