Evaluation of a Point-of-Care Testing Analyzer for Measuring Peripheral Blood Leukocytes
Published on: March 22, 2022
[The investigantion of comparative experiments for different clinical laboratory instruments]
Jia Liu1, Lin Chen, Jing-xia Guo
1Center of Clinical Laboratory, The No. 302 Hospital of the PLA, Beijing 100039, China.
Objective:
To estimate the consistency of two VITROS 3600 chemiluminescent analyzers according to the requirement of ISO15189.
Methods:
Verification tests were made for precision and accuracy of anti-HCV in two instruments. While 40 serum samples including Anti-HCV negative (10 cases) , positive (10 cases) , and weakly positive (20 cases). and the test results were statistical analised.
Results:
Two instruments negative and positive control samples intra-batch precision and coefficients of variation were 5% , 4% and 7. 14% , 7. 23% , inter-batch precision and coefficients of variation were 9. 47% , 7. 7% and 8.04%, 7. 6%, are less than requirement CV (15%) by ISO15189. The accuracy of two instrument were 100% , The test results of the control samples showed no significant difference (P < 0. 05). The correlation analysis of the test results of clinical samples R2 =0. 9984, with good consistency.
Conclusion:
Test results of two Vitros 3600 has good consistency and comparability.
More Related Videos
Related Concept Videos
Automated Microbial Diagnostics
Bioequivalence Experimental Study Designs: Repeated Measures, Cross-Over, Carry-Over, and Latin Square Designs
Data Collection by Experiments
An example of the experimental method is a public clinical trial...
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...