Characteristic variation in evoked potential amplitude with changes in pacing stimulus strength
A B Curtis1, F Vance, K Miller-Shifrin
1Division of Cardiology, University of Florida, Gainesville 32610.
The American Journal of Cardiology
|August 15, 1990
Summary
Evoked potentials, used in cardiac pacing, diminish at higher stimulus amplitudes. This reduction may be due to anodal stimulation from pacing leads, impacting signal analysis for rate-responsive pacing.
Area of Science:
- Biomedical Engineering
- Cardiology
- Electrophysiology
Background:
- Evoked potentials are intracardiac signals from pacing stimuli.
- They show potential for rate-responsive pacing and automatic threshold determination.
- Understanding factors affecting evoked potential morphology is crucial for accurate signal analysis.
Purpose of the Study:
- To investigate the effect of varying stimulus amplitudes on evoked potential morphology.
- To explore the role of anode configuration in evoked potential signal characteristics.
Main Methods:
- Utilized a computer-based pacing system emulator to deliver stimuli (2.5, 5.0, 6.9 V) to 12 patients with bipolar pacing leads.
- Analyzed evoked potential amplitudes and polarity at different voltage settings.
- Evaluated evoked potentials in 8 patients using temporary quadripolar electrode catheters with varied anode configurations.
Main Results:
- Evoked potential amplitude significantly decreased (p < 0.01) and polarity reversed at higher pacing amplitudes (5.0 and 6.9 V) compared to 2.5 V.
- Modifying anode size and current density in quadripolar catheters reduced evoked potential amplitude (p < 0.05).
- High stimulus amplitudes and specific anode configurations diminished the evoked potential signal.
Conclusions:
- Increased stimulus amplitudes reduce evoked potential amplitude and can reverse polarity.
- Anodal stimulation, influenced by pacing lead design, likely contributes to evoked potential signal degradation.
- Findings suggest potential limitations in using evoked potentials for rate-responsive pacing under certain conditions.