Comparison of haemoglobin measurement methods in the operating theatre
B Giraud1, D Frasca, B Debaene
1Service d'Anesthésie et de Réanimation, CHU de Poitiers, Poitiers, France.
Background:
Various methods of haemoglobin (Hb) measurement are available to guide transfusion including several methods that allow for measurement at the bedside. This study directly compared their absolute and trend accuracy compared with values from the central lab (reference method).
Methods:
Adult patients undergoing surgery with expected blood loss wore a rainbow ReSposable sensor connected to a Radical-7 Pulse CO-Oximeter (SpHb). Arterial samples were analysed with a haematology analyser (HbLab), a satellite CO-Oximeter (HbSat), and a point-of-care haemoglobinometer (HemoCue; HcueArt). Concomitantly, ear capillary blood was tested using the same haemoglobinometer (HcueCap). Absolute accuracy and the clinical significance of error were assessed with Bland-Altman plots and three-zone error grids. Trend analysis was performed using a modified polar plot, testing both directionality and magnitude of Hb changes compared with the reference.
Results:
Two hundred and nineteen measurements from 53 patients with HbLab ranging between 6.8 and 16.3 g dl(-1) (4.2 and 10.1 mmol litre(-1)) were recorded. Compared with the reference method, bias (precision) was 0.2 (0.2) g dl(-1) [0.1 (0.1) mmol litre(-1)] for HcueArt, 0.8 (0.3) g dl(-1) [0.5 (0.2) mmol litre(-1)] for HbSat, 0.5 (0.5) g dl(-1) [0.3 (0.3) mmol litre(-1)] for HcueCap and 1.0 (1.2) g dl(-1) [0.6 (0.7) mmol litre(-1)] for SpHb. None of the devices tested would have led to unnecessary or delayed transfusion according to 2006 ASA transfusion criteria. Trend accuracy was better for HcueArt and HbSat than for HcueCap and SpHb.
Conclusion:
Bedside Hb measurement methods differ in their agreement to a laboratory haematology analyser but none would have led to transfusion errors. Trial Registry Number RCB 2009-AO1144-53.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to reabsorb or...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.