Radiation protection in interventional radiology: survey results of attitudes and use
G Emmett Lynskey1, Daniel K Powell, Robert G Dixon
1Department of Radiology, Georgetown University Hospital, Washington, District of Columbia.
Purpose:
To assess attitudes of interventional radiologists toward personal radiation protection and the use of radiation protection devices.
Materials And Methods:
Invitations to an anonymous online survey that comprised eight questions focused on operator attitudes toward radiation protection devices were sent via e-mail to the active membership of the Society of Interventional Radiology (SIR): a total of 3,158 e-mail invitations. A single reminder e-mail was sent.
Results:
There were 504 survey responders (16% response rate). Reported radiation safety device use included lead apron (99%), thyroid shield (94%), leaded eyeglasses (54%), ceiling-suspended leaded shield (44%), rolling leaded shields (12%), ceiling-suspended/rolling lead-equivalent apron (4%), radiation-attenuating sterile surgical gloves (1%), and sterile lead-equivalent patient-mounted drape (4%). Reasons commonly cited for not using certain devices were comfort (eyewear), ease of use (mounted shields), and lack of availability (rolling/hanging shields and patient-mounted shields).
Conclusions:
Interventionalists have an array of tools from which to choose for personal radiation protection; however, for a variety of reasons related to lack of availability or choice, these tools are not universally employed. Further study may be of value to clarify why comfort was cited most often as the primary barrier to the use of protective eyewear and difficulty of use was cited as the primary barrier to use of mounted shields (despite reporting that concern for radiation-induced injury to the eye is paramount). It may also be of interest to further study why certain devices with demonstrable protection effects are not readily available, such as rolling/hanging and patient-mounted shields.
More Related Videos
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
05:18Radiation Planning Assistant - A Web-based Tool to Support High-quality Radiotherapy in Clinics with Limited Resources
Published on: October 6, 2023
Related Concept Videos
Radiation: Applications
The average...
Radiological Investigation I: X-ray and CT
X-ray Imaging
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...