Accurate T(1) mapping for oxygen-enhanced MRI in the mouse lung using a segmented inversion-recovery ultrashort
M Zurek1, E Johansson, F Risse
1AstraZeneca, Personalized Healthcare and Biomarkers, In vivo Biomarkers, Mölndal, Sweden.
Purpose:
A segmented inversion-recovery module combined with the 2D ultrashort echo time radial technique is proposed that allows accurate pixel level T(1) mapping of mouse lung in vivo.
Methods:
Numerical simulations were performed to estimate T(1) measurement accuracy and precision versus flip angle and signal-to-noise ratio. Phantom measurements were used for protocol validation, where the segmented inversion-recovery ultrashort echo-time sequence was compared with the reference technique (inversion-recovery rapid acquisition with refocused echoes). The in vivo experiments were carried out on free-breathing C57 mice (n = 10), breathing first air and then oxygen.
Results:
The simulations demonstrated the high potential of the technique for accurate and precise T(1) assessment. Phantom experiments showed good agreement for T(1) values measured with segmented inversion-recovery ultrashort echo-time and the reference technique. The in vivo experiment demonstrated the utility of the technique in oxygen-enhanced assessment, where small T(1) changes were detected with high precision.
Conclusion:
Segmented inversion-recovery ultrashort echo-time provides accurate, high resolution T(1) mapping of the lung parenchyma.