Imaging In-Stent Restenosis: An Inexpensive, Reliable, and Rapid Preclinical Model
Published on: September 14, 2009
Automatic assessment of stent neointimal coverage by intravascular optical coherence tomography
Giovanni J Ughi1, Christophe J Van Dyck, Tom Adriaenssens
1Department of Cardiovascular Sciences, KU Leuven, Leuven, Belgium.
Aims:
This study aimed to validate automatic intravascular optical coherence tomography (IVOCT) analysis for the evaluation of neointimal coverage in response to stent implantation.
Methods And Results:
Fourteen stented segments in common iliac arteries, acquired from a total of seven adult male New Zealand White rabbits, were interrogated in vivo by IVOCT. Durable polymer everolimus-eluting stents (EES; Xience V, Abbott Vascular, Santa Clara, CA, USA) were used exclusively. Comparison with histology was made in a total of 63 pairs of images, where neointimal thickness over corresponding individual stent struts was assessed. A high correlation coefficient (R = 0.85, P < 0.001) was obtained by comparing automatic IVOCT analysis with histology. Moreover, Bland-Altman statistics showed good limits of agreement (LOAs) of ±45 µm, with an average difference of -10 µm. In addition, manual IVOCT assessment presented very similar results when compared with histology (R = 0.83, P < 0.001 and LOA = ±48 µm with an average difference of -8 µm). Therefore, a very high correlation value was found, comparing manual to automatic IVOCT measurements (R = 0.95, P < 0.001) together with good LOAs (±27 µm) and an average difference of -2 µm.
Conclusion:
The results of the study suggest that automatic IVOCT analysis is a reliable and accurate tool able to speed up current IVOCT analysis procedures. This would potentially allow for a better integration of IVOCT in clinical practice and clinical studies assessing vascular response to stent implantation in a large series of patients.
More Related Videos
13:07Optical Coherence Tomography Based Biomechanical Fluid-Structure Interaction Analysis of Coronary Atherosclerosis Progression
Published on: January 15, 2022
11:12Nondestructive Monitoring of Degradable Scaffold-Based Tissue-Engineered Blood Vessel Development Using Optical Coherence Tomography
Published on: October 3, 2018
Related Concept Videos
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies VII: Vascular Imaging