Prescriptive amplification recommendations for hearing losses with a conductive component and their impact on the
1James H. Quillen VA Medical Center, Mountain Home, TN 37684, USA.
Background:
Hearing aid prescriptive recommendations for hearing losses having a conductive component have received less clinical and research interest than for losses of a sensorineural nature; as a result, much variation remains among current prescriptive methods in their recommendations for conductive and mixed hearing losses (Johnson and Dillon, 2011).
Purpose:
The primary intent of this brief clinical note is to demonstrate differences between two algebraically equivalent expressions of hearing loss, which have been approaches used historically to generate a prescription for hearing losses with a conductive component. When air and bone conduction thresholds are entered into hearing aid prescriptions designed for nonlinear hearing aids, it was hypothesized that that two expressions would not yield equivalent amounts of prescribed insertion gain and output. These differences are examined for their impact on the maximum power output (MPO) requirements of the hearing aid. Subsequently, the MPO capabilities of two common behind-the-ear (BTE) receiver placement alternatives, receiver-in-aid (RIA) and receiver-in-canal (RIC), are examined.
Study Samples:
The two expressions of hearing losses examined were the 25% ABG + AC approach and the 75% ABG + BC approach, where ABG refers to air-bone gap, AC refers to air-conduction threshold, and BC refers to bone-conduction threshold. Example hearing loss cases with a conductive component are sampled for calculations. The MPO capabilities of the BTE receiver placements in commercially-available products were obtained from hearing aids on the U.S. federal purchasing contract.
Results:
Prescribed gain and the required MPO differs markedly between the two approaches. The 75% ABG + BC approach prescribes a compression ratio that is reflective of the amount of sensorineural hearing loss. Not all hearing aids will have the MPO capabilities to support the output requirements for fitting hearing losses with a large conductive component particularly when combined with significant sensorineural hearing loss. Generally, current RIA BTE products have greater output capabilities than RIC BTE products.
Conclusions:
The 75% ABG + BC approach is more appropriate than the 25% ABG + AC approach because the latter approach inappropriately uses AC thresholds as the basis for determining the compression ratio. That is, for hearing losses with a conductive component, the AC thresholds are not a measure of sensorineural hearing loss and cannot serve as the basis for determining the amount of desired compression. The Australian National Acoustic Laboratories has been using the 75% ABG + BC approach in lieu of the 25% ABG + AC approach since its release of the National Acoustic Laboratories--Non-linear 1 (NAL-NL1) prescriptive method in 1999. Future research may examine whether individuals with conductive hearing loss benefit or prefer more than 75% restoration of the conductive component provided adequate MPO capabilities to support such restoration.
Related Concept Videos
Design Example: Vintage Mixing Console
The specifications for the pre-amplifier were clear. It needed to amplify the audio signal by a factor of 10, have an input impedance above 10...
Hearing
Sound Intensity Level
The human ear can perceive an extensive range of sound intensity, necessitating the use of the logarithmic scale to define a physical quantity—the intensity level. It is a ratio of two intensities and hence a...
Maximum Power Transfer
By substituting the entire circuit with...
Perception of Sound Waves
The pitch of a sound depends on the frequency and the pressure amplitude of the source. Two sounds of the same frequency...
The Maximum Power Transfer Theorem
The load connected draws the current, and the circuit delivers the power to the load. The alternating current flowing through the load is determined using the rectangular form of voltages, currents, network impedance, and load impedance. The average power delivered to the load is obtained from the product of the square of current and load resistance.