Evaluation of the particulate concentration in a gelatin-based phantom for sonographically guided lesion biopsy
Jan F Gerstenmaier1, Colin J McCarthy, David P Brophy
1Department of Radiology, St Vincent's University Hospital, Elm Park, Dublin 4, Ireland. gerstenmaierj@fastmail.us
Objectives:
The purpose of this study was to determine the particulate concentration in a gelatin-based ultrasound phantom for lesion biopsy at 6 cm in depth to reduce visualization of the biopsy needle in the near field, simulating subcutaneous fat and tissue echogenicity, and maintain target lesion visualization.
Methods:
Four gelatin-based phantoms with cornstarch at concentrations of 4, 8, 12, and 16 g/L and an anechoic gelatin target at 7 cm in depth were rated on a 5-point scale by readers for visibility of the target lesion, similarity of near-field to abdominal subcutaneous fat echogenicity, and visibility of a 22-gauge spinal needle in the phantom. A timed sonographically guided localization task was performed on the anechoic target by 4 radiology residents using the 22-gauge spinal needle. Results were analyzed by comparative statistical analysis.
Results:
An increasing particulate concentration did not alter the similarity of near-field to abdominal subcutaneous fat echogenicity (P = .6) but did significantly reduce visibility of the anechoic target at a cornstarch concentration of 16 g/L (P = .04) and the 22-gauge needle at 12 g/L (P = .03). Decreased visualization of the needle or target lesion did not affect the time for needle localization of the anechoic target (P = .96).
Conclusions:
The optimal ultrasound phantom cornstarch concentration was 12 g/L to reduce visualization of the spinal needle, simulating subcutaneous fat echogenicity while maintaining target lesion visualization.