Purpose:
Assessment of radiologist's hand dose in CT-guided interventions and determination of influencing factors.
Materials And Methods:
The following CT-guided interventions were included: Core biopsy, drainage, periradicular therapy, and celiac plexus neurolysis. The hand dose was measured with an immediately readable dosimeter, the EDD-30 (Unfors, Sweden). The default parameters for CT fluoroscopy were 120 kV, 90 mA and a 4 mm slice thickness. All interventions were performed on a 16-slice CT unit (Aquilion 16 Toshiba, Japan). The tumor size, degree of difficulty (1 - 3), level of experience and device parameters (mAs, dose-length product, scan time) were documented.
Results:
138 CT-guided interventions (biopsy n = 99, drainage n = 23, pain therapy n = 16) at different locations (lung n = 41, retroperitoneum n = 53, liver n = 25, spine n = 19) were included. The lesion size was 4 - 240 mm (median: 23 mm). The fluoroscopy time per intervention was 4.6 - 140.2 s (median: 24.2 s). The measured hand dose ranged from 0.001 - 3.02 mSv (median: 0.22 mSv). The median hand dose for lung puncture (n = 41) was slightly higher (median: 0.32 mSv, p = 0.01) compared to that for the liver, retroperitoneum and other. Besides physical influencing factors, the degree of difficulty (p = 0.001) and summed puncture depth (p = 0.004) correlated significantly with the hand dose.
Conclusion:
The median hand dose for different CT-guided interventions was 0.22 mSv. Therefore, the annual hand dose limit would normally only be reached with about 2000 interventions.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
08:34Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...