Related Experiment Video
Updated: May 9, 2026

A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy (PRRT): 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Improvements on a patient-specific dose estimation system in nuclear medicine examination
1Department of Biomedical Engineering and Environmental Sciences, National Tsing-Hua University, 101 Section II, Kwang-Fu Rd, Hsin-Chu 30013, Taiwan.
Abstract:
The purpose of this paper is to develop a patient-specific dose estimation system in nuclear medicine examination. A dose deposition routine to store the deposited energy of the photons during their flights was embedded in the widely used SimSET Monte Carlo code and a user-friendly interface for reading PET and CT images was developed. Dose calculated on ORNL phantom was used to validate the accuracy of this system. The ratios of S value for (99m)Tc, (18)F and (131)I computed by this system to those obtained with OLINDA for various organs were ranged from 0.93 to 1.18, which were comparable to that obtained from MCNPX2.6 code (0.88-1.22). Our system developed provides opportunity for tumor dose estimation which cannot be known from the MIRD. The radiation dose can provide useful information in the amount of radioisotopes to be administered in radioimmunotherapy.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
10:33Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification (ADCI) and Dose Estimation
Published on: September 4, 2017
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...