Variable angle locking intercarpal fusion system for four-corner arthrodesis: indications and surgical technique
Juan González Del Pino1, Douglas Campbell, Thomas Fischer
1AO Hand Expert Group (HAEG), Davos, Switzerland.
This study introduces a novel circular plate system for four-corner fusion (4CF) in wrist disorders. The system ensures precise fixation and rapid bone union, aiming to improve outcomes and reduce complications in wrist fusion surgery.
Area of Science:
- Orthopedic Surgery
- Hand and Wrist Surgery
- Biomedical Engineering
Background:
- Four-corner fusion (4CF) is a standard procedure for posttraumatic wrist degeneration.
- Achieving rapid, solid bone union without hardware complications in 4CF remains a challenge.
- Existing implants for 4CF have reported issues, necessitating improved techniques.
Purpose of the Study:
- To present a novel implant system and surgical technique for four-corner fusion (4CF).
- To ensure precise positioning, firm fixation, and solid bone union at the midcarpal joint.
- To facilitate early postoperative recovery and minimize hardware-related complications.
Main Methods:
- Development of a circular plate with variable angle locking and compression screws.
- Design of a reaming-distraction-compression guide for plate positioning and joint preparation.
- Surgical technique utilizing the new implant system for 4CF, including bone grafting.
Main Results:
- The described implant system provides a very solid construct due to locking technology.
- The technique allows for precise fixation and firm attachment to carpal bones.
- The specialized guide facilitates debridement and bone graft interposition for enhanced fusion.
Conclusions:
- The novel circular plate system offers a precise and effective method for four-corner fusion (4CF).
- This system aims to achieve rapid solid bone union and improve patient outcomes in wrist fusion.
- The described technique and implant potentially overcome limitations of previous 4CF hardware.