Dynamic Electrochemical Measurement of Chloride Ions
Published on: February 5, 2016
[Sweat chloride measurement using direct potentiometry: Spotchem(®) (Elitech-Arkray) evaluation and comparison with
Thao Nguyen-Khoa1, Jean-Pierre Borgard, Ryad Miled
1Service de biochimie, Hôpital universitaire Necker-Enfants Malades AP-HP, Paris, France. thao.nguyen-khoa@nck.aphp.fr
Abstract:
Sweat chloride (Cl(-)) measurement is a key step for the diagnosis of cystic fibrosis. The coulometric technique is validated in this context by international guidelines. The aim of our study was to evaluate the assay for sweat Cl(-) ions using direct potentiometry on disposable cassette (Spotchem™ SE EL-1520, Elitech-Arkray) by comparing results to those obtained on the same sample, by coulometry (Chloride analyser Sherwood 926S, Dutscher). To complete our table of correspondence between the results of Cl(-) ions and sweat conductivity (Sweat Check™ 3100), conductivity has been also achieved for 99 of the 139 sweat samples studied. Linearity of each technique performed extends from 10 to 120 mmol/L. The coefficients of variation within and between runs are < 5%. Comparison of 139 results (Passing - Bablock regression) shows a significant difference (p < 0.001): [Spotchem] = 1.026 [Chloride analyser] + 1.8, r = 0.996. After correction with regression factors, only 6 pairs of values (4.6%) had a difference greater than ± 5 mmol/L). The results of conductivity measurement is strongly correlated with those of Cl(-) ions (r = 0.959 for Chloride analyser and 0.965 for Spotchem; p = 0.576) with a linear relationship between the decision thresholds from 30 to 60 mmol/L Cl(-). Sweat Cl(-) determinations using Spotchem™ analyser meet the criteria required by analytical recommendations. The technique is standardized, easy to perform and fast. Its good practicability makes the sweat test independent to operator and allows point-of care use.
Related Concept Videos
Controlled-Potential Coulometry: Electrolytic Methods
The chosen potential ensures...
Controlled-Current Coulometry: Overview
Potentiometry: Overview
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...
Potentiometry: Membrane Electrodes
Coulometry: Overview