Diagnostic accuracy of laser Doppler flowmetry versus strain gauge plethysmography for segmental pressure measurement
Christian Høyer1, Jes Sandermann, Jens Peder D Paludan
1Department of Clinical Physiology, Viborg Regional Hospital, Viborg, Denmark; Department of Health Science and Technology, Aalborg University, Aalborg, Denmark.
Objective:
To assess the diagnostic accuracy of laser Doppler flowmetry (LDF) with mercury-in-silastic strain gauge plethysmography (SGP) as a reference test for measuring the toe and ankle pressures in patients with known or suspected peripheral arterial disease (PAD).
Methods:
This was a prospective, randomized, blinded diagnostic accuracy study. Toe and ankle pressures were measured using both methods in 200 consecutive patients, who were recruited at our vascular laboratory over a period of 30 working days. Classification of PAD and critical limb ischemia (CLI) was made in accordance with TASC-II criteria.
Results:
The LDF method demonstrated 5.8 mm Hg higher mean toe pressures than the SGP method for the right limb and 7.0 mm Hg for the left limb (both P < .001). There were no significant differences in the mean ankle pressures (both P > .129). The limits of agreement for the differences (SGP - LDF) were -31.7 to 20.2 mm Hg for right toe pressures, -28.0 to 14.0 mm Hg for left toe pressures, -25.5 to 22.8 mm Hg for right ankle pressures, and -26.9 to 24.6 mm Hg for left ankle pressures. A correlation analysis of the absolute pressures using the two methods showed an intraclass correlation coefficient of 0.902 (95% confidence interval [CI], 0.835-0.938) for right toe pressures, 0.919 (95% CI, 0.782-0.960) for the left toe pressures, 0.953 (95% CI, 0.937-0.965) for right ankle pressures, and 0.952 (95% CI, 0.936-0.964) for left ankle pressures. Cohen's Kappa showed an agreement in the diagnostic classification of κ = 0.775 (95% CI, 0.631-0.919) for PAD and κ = 0.780 (95% CI, 0.624-0.936) for CLI.
Conclusions:
LDF showed a good correlation with SGP over a wide range of toe and ankle pressures, as well as substantial agreement for the diagnostic classification of PAD including CLI.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Peripheral Arterial Disease II: Clinical Manifestations and Diagnostic Evaluation
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Pipe Flowrate Measurement
The orifice meter is a simple,...