Related Experiment Video
Updated: May 9, 2026

Field-Based Thermal Physiology Assay: Cold Shock Recovery under Ambient Conditions
Published on: March 9, 2021
The heat is on: room temperature affects laboratory equipment--an observational study
Julia M Butler1, Jane E Johnson, William R Boone
1Department of Obstetrics and Gynecology, Greenville Health System University Medical Group, 890 W. Faris Rd., Suite 470, Greenville, SC, USA, jbutler@ghs.org.
Objective:
To evaluate the effect of ambient room temperature on equipment typically used in in vitro fertilization (IVF).
Design:
We set the control temperature of the room to 20 °C (+/-0.3) and used CIMScan probes to record temperatures of the following equipment: six microscope heating stages, four incubators, five slide warmers and three heating blocks. We then increased the room temperature to 26 °C (+/-0.3) or decreased it to 17 °C (+/-0.3) and monitored the same equipment again. We wanted to determine what role, if any, changing room temperature has on equipment temperature fluctuation.
Results:
There was a direct relationship between room temperature and equipment temperature stability. When room temperature increased or decreased, equipment temperature reacted in a corresponding manner. Statistical differences between equipment were found when the room temperature changed. What is also noteworthy is that temperature of equipment responded within 5 min to a change in room temperature.
Conclusions:
Clearly, it is necessary to be aware of the affect of room temperature on equipment when performing assisted reproductive procedures. Room and equipment temperatures should be monitored faithfully and adjusted as frequently as needed, so that consistent culture conditions can be maintained. If more stringent temperature control can be achieved, human assisted reproduction success rates may improve.
Related Concept Videos
Physical Methods for Controlling Microbial Growth: Temperature
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Effect of Temperature Change on Reaction Rate
Factors Affecting Body Temperature
Factors may include:
Effects of Temperature on Free Energy
Factors Influencing Microbial Growth: Temperature