In vitro evaluation of positive expiratory pressure devices attached to nebulizers
1Pulmonology Section, Department of Pediatrics, University of Arkansas for Medical Sciences, and with the Pediatric Aerosol Research Laboratory, Arkansas Children's Hospital Research Institute, Little Rock, Arkansas.
Background:
Patients with cystic fibrosis perform airway clearance techniques and receive nebulized medications on a regular basis. Some positive expiratory pressure (PEP) devices allow concomitant administration of aerosol. I hypothesized that this practice alters the aerosol characteristics and patient dose. I compared the aerosol characteristics and patient dose of nebulized albuterol from 2 types of nebulizer, alone and when connected to different PEP and vibratory PEP devices.
Methods:
Three units of a continuous-output nebulizer (Up-Draft II Opti-Neb) and 3 units of a breath-enhanced nebulizer (LC Plus) were tested alone and connected to PEP devices (Acapella Choice, Acapella Duet, and EzPAP for Up-Draft II Opti-Neb, and Pari PEP at 2 different settings, and Pari PEP S system with the LC Plus). Aerosol characteristics were evaluated with a cooled cascade impaction technique. The nebulizers were loaded with 2.5 mg/3 mL albuterol solution and operated for 4 min at 6 L/min (wall air). Patient dose was evaluated with simulated breathing patterns for a child, small adult, and large adult. Albuterol was assayed via spectrophotometry.
Results:
Connecting the LC Plus to the PEP devices did not change the aerosol characteristics or patient dose. Connecting the Up-Draft II Opti-Neb to the PEP devices significantly reduced the mass median aerodynamic diameter, from 4.13 μm to 3.72 μm with EzPAP (P = .02), 1.24 μm with Acapella Choice (P < .001), and 1.22 μm with Acapella Duet (P < .001). The total amount of albuterol captured by the impactor decreased when connected to either Acapella Choice (65%) or Acapella Duet (69%), with 17-25% retained in the PEP devices. Patient dose decreased by 76% and 84% when connected to Acapella Choice and Acapella Duet, respectively.
Conclusions:
Concomitant use of nebulizer and PEP or vibratory PEP devices that obstruct the aerosol pathway significantly decrease the aerosol particle size and the patient dose.
More Related Videos
07:28Dry Powder and Nebulized Aerosol Inhalation of Pharmaceuticals Delivered to Mice Using a Nose-only Exposure System
Published on: April 6, 2017
05:46A Novel Inhalation Mask System to Deliver High Concentrations of Nitric Oxide Gas in Spontaneously Breathing Subjects
Published on: May 4, 2021
Related Concept Videos
Pulmonary Function Tests
Pulmonary Function Tests are crucial diagnostic tools for assessing respiratory function, particularly in patients with chronic respiratory disorders. They comprehensively evaluate lung volumes, ventilatory function, breathing mechanics, diffusion, and gas exchange. These tests help diagnose pulmonary diseases and play a significant role in monitoring disease progression, evaluating disability, and assessing response to therapy.
PFTs involve using a spirometer, a...
Inhaled Medications
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...