Background:
- Total hip arthroplasty (THA) aims to restore function and alleviate pain in patients with hip joint degeneration.
- Traditional THA implants often involve significant bone stock removal, potentially impacting long-term outcomes.
- The Thrust Plate Prosthesis is designed to align with physiologic loading principles of the proximal femur, aiming to preserve bone stock.
Purpose of the Study:
- To evaluate the long-term clinical and radiological outcomes of the Thrust Plate Prosthesis in total hip arthroplasty.
- To determine the reoperation-free survival rate of this implant.
- To assess patient-reported pain and radiographic findings associated with the Thrust Plate Prosthesis.
Main Methods:
- Retrospective analysis of 34 patients (36 prostheses) who underwent THA with the Thrust Plate Prosthesis.
- Investigation of reoperation-free survival, clinical symptoms (pain), and radiological parameters (radiolucencies, bone resorption, hypertrophy).
- Mean follow-up duration of 11.9 years in patients with a mean age of 51 years at operation.
Main Results:
- Reoperation-free survival was 88.9% over the follow-up period.
- Four reoperations (11.1%) were performed: three for aseptic loosening and one for a distal fracture.
- Common clinical complaint was lateral hip pain (44%); radiological findings included minimal radiolucencies (<1 mm) and significant calcar hypertrophy (97%).
Conclusions:
- The Thrust Plate Prosthesis demonstrates a relatively high reoperation-free survival rate and favorable radiological results in long-term follow-up.
- It presents a potential alternative to traditional stemmed total hip arthroplasty, particularly for younger patient populations.