Related Experiment Video
Updated: May 9, 2026

Assessing Early Stage Open-Angle Glaucoma in Patients by Isolated-Check Visual Evoked Potential
Published on: May 25, 2020
Screening for glaucoma using GDx-VCC in a population with ≥1 risk factors
Gisèle Li1, Alvine Kamdeu Fansi, Paul Harasymowycz
1Ophthalmology Research Unit, Maisonneuve-Rosemont Hospital, University of Montreal, Montreal, Que., Canada. gisele.li@umontreal.ca
The GDx-VCC glaucoma screening tool showed low sensitivity for detecting definitive glaucoma, despite high specificity. Further research is needed to improve its accuracy in at-risk populations.
Area of Science:
- Ophthalmology
- Medical Diagnostics
- Public Health
Background:
- Glaucoma is a leading cause of irreversible blindness worldwide.
- Early detection through screening is crucial for managing glaucoma and preventing vision loss.
- The GDx-VCC is a non-invasive device used for assessing retinal nerve fiber layer (RNFL) thickness, a key indicator of glaucoma.
Purpose of the Study:
- To evaluate the diagnostic accuracy of the GDx-VCC for screening glaucoma in individuals with at least one risk factor.
- To determine the sensitivity, specificity, and predictive values of the GDx-VCC in a high-risk population.
Main Methods:
- A cross-sectional study design was employed to assess the GDx-VCC's performance.
- Two hundred forty-seven community volunteers with glaucoma risk factors participated.
- Retinal nerve fiber layer (RNFL) thickness was measured using GDx-VCC, with glaucoma classification based on ophthalmologic examination and frequency doubling perimetry.
Main Results:
- Data from 219 eyes were analyzed after exclusions for poor scan quality or missing data.
- The nerve fiber indicator parameter, with a cutoff of 35, demonstrated 75% sensitivity and 95% specificity.
- Positive predictive value was 25%, and negative predictive value was 99%, with a positive likelihood ratio of 16.
Conclusions:
- The GDx-VCC exhibits inadequate sensitivity for the effective screening of definitive glaucoma.
- While specificity is high, the low sensitivity limits its utility as a standalone screening tool for glaucoma detection in at-risk individuals.
Related Concept Videos
Glaucoma: Overview
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Angle Closure Glaucoma: Treatment
Diabetic Retinopathy
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT