Effective dose to staff members in a positron emission tomography/CT facility using zirconium-89
1Department of Radiological Sciences, King Saud University, Riyadh, Saudi Arabia. kalzimami@ksu.edu.sa
Objective:
Positron emission tomography (PET) using zirconium-89 ((89)Zr) is complicated by its complex decay scheme. In this study, we quantified the effective dose from (89)Zr and compared it with fluorine-18 fludeoxyglucose ((18)F-FDG).
Methods:
Effective dose distribution in a PET/CT facility in Riyadh was calculated by Monte Carlo simulations using MCNPX. The positron bremsstrahlung, the annihilation photons, the delayed gammas from (89)Zr and those emissions from (18)F-FDG were modelled in the simulations but low-energy characteristic X-rays were ignored.
Results:
On the basis of injected activity, the dose from (89)Zr was higher than that of (18)F-FDG. However, the dose per scan from (89)Zr became less than that from (18)F-FDG near the patient, owing to the difference in injected activities. In the corridor and control rooms, the (89)Zr dose was much higher than (18)F-FDG, owing to the difference in attenuation by the shielding materials.
Conclusion:
The presence of the high-energy photons from (89)Zr-labelled immuno-PET radiopharmaceuticals causes a significantly higher effective dose than (18)F-FDG to the staff outside the patient room. Conversely, despite the low administered activity of (89)Zr, it gives rise to a comparable or even lower dose than (18)F-FDG to the staff near the patient. This interesting result raises apparently contradictory implications in the radiation protection considerations of a PET/CT facility.
Advances In Knowledge:
To the best of our knowledge, radiation exposure to staff and public in the PET/CT unit using (89)Zr has not been investigated. The ultimate output of this study will lead to the optimal design of the facility for routine use of (89)Zr.
More Related Videos
09:03Radiotracer Administration for High Temporal Resolution Positron Emission Tomography of the Human Brain: Application to FDG-fPET
Published on: October 22, 2019
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy (PRRT): 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation I: X-ray and CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Biological Effects of Radiation
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...