Related Experiment Video
Updated: May 8, 2026

A Computerized Functional Skills Assessment and Training Program Targeting Technology Based Everyday Functional Skills
Published on: February 13, 2020
Safety-oriented usability test of a semi-automated unit dose system: role of task allocation between human and
Sylvia Pelayo1, Sylvain Hassler, Stéphanie Bernonville
1INSERM CIC-IT, EVALAB, Lille; CHU Lille; Univ Lille Nord de France, UDSL EA 2694; F-59000 Lille, France.
Abstract:
The distribution of tasks between humans and machines in the design of healthcare systems is an important issue for patient safety. This paper presents a usability test performed to compare a semi-automated unit dose system (UDS) with the usual/manual preparation procedure for preparing and administering drugs. The results show that the UDS prevents the frequent administration errors encountered with the usual cabinets and produces a better performance in terms of time for filling the pill dispensers (6.52 sec ± 1.1 vs. 8.5 sec ± 1.5 (t(9, 16) = 3.12, p <.007)). But the results also stress that the UDS takes entire control of the preparation and administration tasks, thus leading to a loss of control of the process by nurses (difficulties experienced by them in resuming their actions, difficulties in memorizing drugs, lack of confidence in the UDS while they "blindly" rely on it). The distribution of tasks between the nurses and the UDS should be modified to give back control to the nurses. Design suggestions were provided in this way. For instance, the UDS may guide the nurses for the drugs localization in the cabinet but leaves the validation of the drugs to the nurses.