Evaluation of postprocessing dual-energy methods in quantitative computed tomography. Part 2. Practical aspects.

C van Kuijk1, J L Grashuis, J C Steenbeek

  • 1Department of Diagnostic Radiology, Erasmus University Rotterdam/Academic Hospital Rotterdam, The Netherlands.

Investigative Radiology
|August 1, 1990
PubMed
Summary

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Incremental diagnostic value of epicardial adipose tissue for the detection of functionally relevant coronary artery disease.

Atherosclerosis·2015
Same author

The impact of obesity on the relationship between epicardial adipose tissue, left ventricular mass and coronary microvascular function.

European journal of nuclear medicine and molecular imaging·2015
Same author

Anaesthetic management during open and percutaneous irreversible electroporation.

British journal of anaesthesia·2014
Same author

Ablation of colorectal liver metastases by irreversible electroporation: results of the COLDFIRE-I ablate-and-resect study.

European radiology·2014
Same author

Transcatheter CT arterial portography and CT hepatic arteriography for liver tumor visualization during percutaneous ablation.

Journal of vascular and interventional radiology : JVIR·2014
Same author

MR imaging of bladder endometriosis and its relationship with the anterior uterine wall: experience in a tertiary referral centre.

European journal of radiology·2011

Dual-energy quantitative computed tomography (DEQCT) calibration accuracy depends on material choice and positioning. Central calibration and using materials mimicking bone components yield the best results for DEQCT.

Area of Science:

  • Medical Imaging
  • Radiology
  • Quantitative Computed Tomography

Background:

  • Dual-energy quantitative computed tomography (DEQCT) enables material decomposition.
  • Accurate DEQCT requires precise calibration using tissue-equivalent materials.
  • Understanding calibration variables is crucial for reliable DEQCT analysis.

Purpose of the Study:

  • To evaluate DEQCT postprocessing algorithms.
  • To assess the impact of tissue-equivalent material selection on calibration accuracy.
  • To compare central versus peripheral calibration methods in DEQCT.

Main Methods:

  • Postprocessing algorithms evaluated: Cann, Laval-Jeantet et al, Goodsitt et al (2 methods), and Nickoloff et al.
  • Tissue-equivalent materials tested: bone mineral (K2HPO4, hydroxyapatite), fat (paraffin, polyethylene, ethanol), and red marrow (plastic).

Related Experiment Videos

  • Calibration methods compared: central and peripheral positioning of materials.
  • Main Results:

    • Central calibration positioning minimized deviation from manufacturer's quoted content.
    • Using identical tissue-equivalent materials for calibration and measurement improved accuracy.
    • Goodsitt et al (2 methods) and Nickoloff et al algorithms demonstrated superior performance.
    • Material choice significantly impacts DEQCT accuracy, emphasizing the need for bone-mimicking materials.

    Conclusions:

    • The selection of appropriate tissue-equivalent materials is critical for accurate DEQCT.
    • Central calibration and algorithm choice (Goodsitt et al, Nickoloff et al) enhance DEQCT reliability.
    • Future DEQCT applications should prioritize materials closely resembling bone composition for precise quantitative analysis.