Related Experiment Video
Updated: May 8, 2026

Comparative Analysis of Human Growth Hormone in Serum Using SPRi, Nano-SPRi and ELISA Assays
Published on: January 7, 2016
Implementation of a SPR immunosensor for the simultaneous detection of the 22K and 20K hGH isoforms in human serum
Elena de Juan-Franco1, J M Rodríguez-Frade, M Mellado
1Nanobiosensors and Bioanalytical Applications Group, Research Center on Nanoscience and Nanotechnology (CSIC) and CIBER-BBN, 08193 Bellaterra, Barcelona, Spain. edjuanf@gmail.com
Abstract:
We have implemented a Surface Plasmon Resonance (SPR) immunosensor based on a sandwich assay for the simultaneous detection of the two main hGH isoforms, of 22 kDa (22K) and 20 kDa (20K). An oriented-antibody sensor surface specific for both hormone isoforms was assembled by using the biotin-streptavidin system. The immunosensor functionality was checked for the direct detection of the 22K hGH isoform in buffer, which gave high specificity and reproducibility (intra and inter-assay mean coefficients of variation of 8.23% and 9% respectively). The selective determination of the 22K and 20K hGH isoforms in human serum samples in a single assay was possible by using two specific anti-hGH monoclonal antibodies. The detection limit for both hormone isoforms was 0.9 ng mL(-1) and the mean coefficient of variation was below 7.2%. The excellent reproducibility and sensitivity obtained indicate the high performance of this immunosensor for implementing an anti-doping test.
More Related Videos
08:28Validated Immunochemical Assay for Comprehensive Determination of the Human Epidermal Growth Factor Receptor 2 Released from and Bound to Cells
Published on: May 9, 2025
08:26Development and Validation of an Ultrasensitive Single Molecule Array Digital Enzyme-linked Immunosorbent Assay for Human Interferon-α
Published on: June 14, 2018