Objectives:
The EZ-Blocker (IQ Medical Ventures BV, Rotterdam, Netherlands) is a newly designed device for one-lung ventilation. The aim of this study was to compare the effectiveness of the Cohen Flex-Tip bronchial blocker (Cook, Bloomington, IN) and the EZ-Blocker for one-lung ventilation during thoracic surgery.
Design:
Randomized and prospective.
Setting:
A university hospital.
Participants:
This study included 40 patients undergoing thoracic surgical procedures.
Interventions:
Patients were assigned to 2 study groups: Patients who received the Cohen Flex-Tip blocker were assigned to the Cohen group, and patients who received the EZ-Blocker were assigned to the EZ group. In both groups, fiberoptic guidance was used during placement of the bronchial blockers. Comparisons between the groups included the time to correct placement, the incidence of malpositioning, and the satisfaction level of the surgeon (good, fair, poor).
Measurements And Main Results:
One-lung ventilation was achieved successfully for all patients. The time to correct placement (mean±SD) was significantly shorter in the EZ group (146±56 seconds) compared with the Cohen group (241±51 seconds; p=0.01). The incidence of malpositioning was significantly lower in the EZ group compared with the Cohen group (p=0.018). Surgeon satisfaction was similar in both groups.
Conclusions:
In this study, both bronchial blockers provided similar surgical exposure during thoracic procedures. The EZ-Blocker had a shorter time to correct positioning and less frequent intraoperative malpositioning.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Cardiopulmonary Resuscitation II: ACLS Airway Management