Influence of face mask design on bag-valve-mask ventilation performance: a randomized simulation study
1Department of Emergency Medicine, Kangbuk Samsung Hospital, Sungkyunkwan University School of Medicine, Seoul, Korea.
Background:
Different face mask designs can influence bag-valve-mask (BVM) ventilation performance during resuscitation. We compared a single-use, air-cushioned face mask (AM) with a reusable silicone face mask (SM) for quality of BVM ventilation on a manikin simulating cardiac arrest.
Methods:
Thirty-two physicians were recruited, and a prospective, randomized, crossover observational study was conducted after an American Heart Association-accredited basic life support provider course and standardized practice time were completed. Participants performed 12 cycles of BVM ventilation with both the AM and SM on a SmartMan lung simulator.
Results:
Mean tidal volume was significantly higher in ventilations performed using the AM vs. the SM (548 ± 159 ml vs. 439 ± 163 ml, P < 0.01). In addition, the proportion of low-volume ventilation was significantly lower with the AM than the SM [6/12 (2-11) vs. 9/12 (5-12), P = 0.03]. Bag-valve-AM ventilation volume was not affected by the physical characteristics of the rescuers, except for sex. In contrast, bag-valve-SM ventilation volume was affected by most of the characteristics tested, including sex, height, weight, hand width, hand length, and grip power.
Conclusion:
The AM seems to be a more efficient face mask than the SM at delivering sufficient ventilation volumes. The performance of the AM did not seem to be associated with the physical characteristics of the rescuers, whereas that of the SM was affected by these factors. The SM may not be an appropriate face mask for performing one-person BVM ventilation during resuscitation for rescuers who are smaller in stature, have a smaller hand size, or have weaker grip power.
Related Concept Videos
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...