Low-contrast-dose protocol in cardiac CT: 20% contrast dose reduction using 100 kVp and high-tube-current-time
Masafumi Kidoh1, Takeshi Nakaura2, Shinichi Nakamura2
1Diagnostic Radiology, Amakusa Medical Center, Kumamoto, Japan Department of Diagnostic Radiology, Graduate School of Medical Sciences, Kumamoto University, Japan masafkidoh@yahoo.co.jp.
Background:
The use of the smallest contrast dose is highly desirable in performing cardiac computed tomography (CT), especially for patients with cardiovascular diseases to prevent contrast-induced nephropathy.
Purpose:
To evaluate the feasibility of 20% reduced contrast dose protocol in cardiac CT using 100 kVp and high-tube-current-time product setting.
Material And Methods:
Fifty patients were scanned with our conventional 120 kVp protocol, and the other 50 patients underwent scans using a tube voltage of 100 kVp, a high-tube-current-time product, and a 20% reduced contrast dose. We evaluated estimated effective dose (ED), CT attenuation, image noise, and contrast-to-noise ratio (CNR) of the ascending aorta. We also evaluated CT attenuation of the coronary arteries. Two radiologists independently assessed image quality of coronary arteries.
Results:
There was no significant difference in the ED between the 100 kVp and 120 kVp protocols (21.7 mSv ± 1.6 vs. 21.8 mSv ± 1.1, P=0.65). There was no significant difference in the CNR of the ascending aorta between the 100 kVp and 120 kVp protocols (18.8 ± 3.5 vs. 18.7 ± 3.8, P=0.98). Mean CT attenuation of the coronary arteries of the 100 kVp protocols was significantly higher than that of 120 kVp protocols (P<0.05). There was no significant difference in the overall image quality of the coronary artery between the 100 kVp and 120 kVp protocols (3.7 ± 0.4 vs. 3.7 ± 0.5, P=0.65).
Conclusion:
For cardiac CT a voltage setting of 100 kVp and a high-tube-current-time product enable 20% reduction in the contrast dose without affecting the quality of coronary artery images compared with a 120 kVp and standard-contrast-dose CT protocol.
More Related Videos
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation I: X-ray and CT
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Cardiac Catheterization I: Pre-Procedure Overview