The microcirculation image quality score: development and preliminary evaluation of a proposed approach to grading

Michael J Massey1, Ethan Larochelle, Gabriel Najarro

  • 1Department of Emergency Medicine, Beth Israel Deaconess Medical Center, Boston, MA.

Journal of Critical Care
|August 27, 2013
PubMed
Abstract

Insights

A new scoring system, the Microcirculation Image Quality Score, objectively assesses bedside sublingual microcirculation videos. This method ensures adequate image quality for analysis, improving diagnostic reliability.

Area of Science:

  • Medical Imaging
  • Microcirculation Research
  • Point-of-Care Diagnostics

Background:

  • Side-stream dark-field microscopy enables direct bedside visualization of sublingual microcirculation.
  • Technical challenges in image acquisition can compromise the quality and reliability of these videos.
  • Standardized quality assessment is needed to ensure accurate analysis of microcirculation data.

Purpose of the Study:

  • To present and assess a novel quality assurance method for rating image acquisition quality in sublingual side-stream dark-field microscopy videos.
  • To establish objective criteria for evaluating video quality before performing microcirculation analysis.
  • To improve the consistency and validity of bedside microcirculation imaging.

Main Methods:

  • A "Microcirculation Image Quality Score" was developed, evaluating 6 key areas: illumination, duration, focus, content, stability, and pressure.
  • Each category was scored as optimal (0 points), acceptable (1 point), or unacceptable (10 points).
  • 59 videos from 34 emergency department patients were assessed by two blinded raters; a cumulative score of 10 or higher indicated unacceptable quality.

Main Results:

  • The Microcirculation Image Quality Score was applied to 59 videos, with mean passing and failing scores of 1.68 and 15.74, respectively.
  • 27 of 59 videos met the quality criteria (score < 10), with pressure artifact being the most common failure point.
  • High interrater agreement (Cohen κ ranging from 0.66 to 1.0) was observed across all categories and overall pass-fail rates.

Conclusions:

  • The developed Microcirculation Image Quality Score effectively addresses common video quality degradation issues.
  • The criteria provide an objective method for assessing image acquisition quality, ensuring adequate video selection for analysis.
  • Preliminary results indicate good interrater reliability, suggesting the score is repeatable and warrants further validation.