Related Experiment Video
Updated: May 8, 2026

Rapid Nanoprobe Signal Enhancement by In Situ Gold Nanoparticle Synthesis
Published on: March 7, 2018
Oxygen biosensor based on bilirubin oxidase immobilized on a nanostructured gold electrode
Marcos Pita1, Cristina Gutierrez-Sanchez, Miguel D Toscano
1Instituto de Catalisis y Petroleoquimica, Consejo Superior de Investigaciones Cientificas, C/Marie Curie 2, L10, 28049 Madrid, Spain.
Abstract:
Gold disk electrodes modified with gold nanoparticles have been used as a scaffold for the covalent immobilization of bilirubin oxidase. The nanostructured bioelectrodes were tested as mediator-less biosensors for oxygen in a buffer that mimics the content and the composition of human physiological fluids. Chronoamperometry measurements showed a detection limit towards oxygen of 6 ± 1 μM with a linear range of 6-300 μM, i.e. exceeding usual physiological ranges of oxygen in human tissues and fluids. The biosensor presented is the first ever-reported oxygen amperometric biosensor based on direct electron transfer of bilirubin oxidase.