Electronic collimation and radiation protection in paediatric digital radiography: revival of the silver lining
J Bomer1, L Wiersma-Deijl, H C Holscher
1Radiology Department, Haga Teaching Hospital/Juliana Children's Hospital, Leyweg 275, 2545 CH, Den Haag, The Netherlands, j.bomer@hagaziekenhuis.nl.
Abstract:
In digital radiography we are now able to electronically collimate images after acquisition. This may seem convenient in paediatric imaging, but we have to be aware that electronic collimation has two major downsides. Electronic collimation implicates that the original field size should have been smaller and the child has been exposed to unnecessary radiation. Also, by use of electronic collimation, potentially important information may be lost. The "silver lining", denoting the X-ray beam collimation, can serve as a useful radiation protection instrument to check for proper field size and detect unnecessary exposure. Furthermore, the silver lining confirms all exposed anatomy is shown in the final image, and thus may also serve as a quality assurance instrument as the patient has the right to all acquired information. Teaching Points • The ability to electronically collimate an image after acquisition may serve to enhance contrast in the region of interest. • The ability to electronically collimate an image after acquisition carries the risk of overexposure. • The ability to electronically collimate an image after acquisition carries the risk of losing important information. • The silver lining can serve as a quality control instrument for proper collimation. • The patient has the right to all information obtained during an X-ray examination.
Related Concept Videos
X-ray Imaging
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...