Evolution of hook design for fixation of the titanium Greenfield filter
L J Greenfield1, K J Cho, J R Tauscher
1Department of Surgery, University of Michigan Hospital, Ann Arbor 48109.
Abstract:
The titanium Greenfield filter was developed to facilitate operative and percutaneous insertion by a carrier (12F) smaller than that required for the standard stainless steel filter (24F). In preliminary clinical studies however, it proved to have an unacceptable rate of distal slippage and caval wall penetration. Therefore four modifications in hook design and base diameter have been tested in 20 sheep by implantation above and below the renal veins. In comparison to control titanium Greenfield filters (n = 8), neither changes in hook angle with smaller diameter base (type I, n = 4), elimination of limb flare (type II, n = 4), or 80 degree hook angle (type IV, n = 10) prevented migration and entry into venous tributaries or penetration. Penetration scores on the basis of autopsy findings did not correlate with antemortem predictions based on cavograms and showed 1.5 +/- 2.0/filter for controls, 1.0 +/- 1.15/filter for type I, 1.75 +/- 0.96/filter for type II and 1.67 +/- 1.37/filter for type IV. It was necessary to change the configuration of the hook completely to a recurved hook design providing a base pad for the 80 degree hook (type VI, n = 6) to limit migration and penetration with a score of 0.33 +/- 0.82/filter. Effects of experimental embolism were studied in six sheep (two acute, four chronic) and showed effective trapping by both type IV and type VI filters with no evidence of filter migration or damage to the caval wall. Persistent thrombus appeared to be caused by the remnants of oxidized cellulose used to induce thrombi, but there was no evidence of propagation proximal to the filter.(ABSTRACT TRUNCATED AT 250 WORDS)