Fluoride uptake by human tooth enamel: topical application versus combined dielectrophoresis and AC electroosmosis
Chris S Ivanoff1, Bashir I Morshed, Timothy L Hottel
1Department of Bioscience Research, College of Dentistry, University of Tennessee Health Science Center, 875 Union Avenue, Memphis, TN 38163, USA. civanoff@uthsc.edu
Purpose:
To compare fluoride uptake by enamel after applying 1.23% acidulated phosphate fluoride gel to human tooth enamel topically (n = 12) or with combined dielectrophoresis and AC electroosmosis (DEP/ACE) at frequencies of 10, 400 and 5,000 Hz (n = 12) for 20 minutes.
Methods:
DEP/ACE induced nonuniform electrical fields with three alternating current frequencies to polarize, orient, and motivate fluoride particles. Fluoride concentrations were measured at various enamel depths using wavelength dispersive spectrometry. Data were analyzed by ANOVA/Student-Newman-Keuls post hoc tests (P < or = 0.05).
Results:
Fluoride concentrations in the diffusion group were significantly higher than baseline readings at 10, 20 and 50 microm depths. Fluoride concentrations in DEP/ACE-treated teeth were significantly higher than the diffusion group at 10, 20, 50, 100, 200 and 300 microm (ANOVA/Student-Newman-Keuls post hoc, P < 0.05). Fluoride uptake with DEP/ACE was substantially higher than diffusion at 10, 20, 50, 100, 200 and 300 microm depths (paired t test, P < 0.05). DEP/ACE transported fluoride up to 300 microm deep, whereas conventional fluoride application was comparatively ineffective beyond 20 microm depth (P < 0.05). Compared to passive diffusion, fluoride uptake in enamel was significantly higher in the DEP/ACE group at 10, 20, 50, 100, 200 and 300 microm depths (P < 0.05). DEP/ACE drove fluoride substantially deeper into human enamel with a difference in uptake 1,575 ppm higher than diffusion at 100 microm depth; 6 times higher at 50 microm depth; 5 times higher at 20 microm depth; and 7 times higher at 10 microm depth. Fluoride levels at 100 microm were equivalent to long-term prophylactic exposure.
Related Concept Videos
Interfacial Electrochemical Methods: Overview
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a result, EDTA...