Related Experiment Video
Updated: May 8, 2026

Arterial Pouch Microsurgical Bifurcation Aneurysm Model in the Rabbit
Published on: May 14, 2020
Endovascular treatment with flow diverters may fail to occlude experimental bifurcation aneurysms
Jean Raymond1, Tim E Darsaut, Alina Makoyeva
1Department of Radiology, Notre-Dame Hospital, Centre Hospitalier de l'Université de Montréal, 1560 Sherbrooke East, Montreal, QC, H2L 4M1, Canada, jean.raymond@umontreal.ca.
Introduction:
Animal models may explain how stents and flow diverters (FDs) may succeed or fail to treat bifurcation aneurysms.
Methods:
In vitro studies were designed to anticipate device deformations in bifurcations. Large, wide-necked bifurcation aneurysms were constructed in 21 animals and treated 4-8 weeks later using stents and FDs in various combinations, forming four main groups: parent artery to right branch flow diversion (RBFD, n = 6), parent artery to left branch flow diversion (LBFD, n = 6), Y flow diversion (YFD, n = 4), and compared with high-porosity Y-stenting (YHPS, n = 4). The results include immediate and follow-up angiography at 3 months, followed by grading of the extent of neointimal coverage of devices at pathology.
Results:
In vitro, all braided devices showed varying porosities according to characteristic zones. FDs can be compacted to decrease porosities, but a limiting factor is the constant presence of a more porous transition zone. In vivo, 3/6 RBFD, 4/6 LBFD, and 2/4 YFD treated aneurysms had decreased in size by 3 months, while those treated with YHPS increased in size (P = 0.15). There was a significant correlation between device porosities and extent of neointimal coverage (r = 0.639, P = 0.002), and between porosities and angiographic evolution (r = -0.655, P = 0.002), but not between neointima formation and angiographic evolution (r = -0.278, P = 0.235). Failures could be explained in all cases by the presence of leaks or holes in the neointima at the level of the transition zones.
Conclusion:
FDs did not successfully treat most bifurcation aneurysms, at least in this animal model.
More Related Videos
11:00Experimental Investigation of Secondary Flow Structures Downstream of a Model Type IV Stent Failure in a 180° Curved Artery Test Section
Published on: July 19, 2016
07:21Microsurgical Creation of Giant Bifurcation Aneurysms in Rabbits for the Evaluation of Endovascular Devices
Published on: September 8, 2023
Related Concept Videos
Aneurysm III: Interprofessional Care
Aneurysm I: Introduction