Related Experiment Videos
Oxford phase III meniscal bearing fracture: case report
Hong-Chul Lim1, Won-Yong Shon1, Seung-Ju Kim1
1Department of Orthopaedic Surgery, Korea University Guro Hospital, Seoul, Republic of Korea.
The Knee
|September 3, 2013
Abstract:
Meniscal bearing fracture is a rare complication of phase III Oxford unicompartmental knee replacement (UKR). We report a case of a meniscal bearing fracture that occurred 7 years after phase III Oxford medial UKR. The meniscal bearing showed uneven delamination of the polyethylene in the thinnest articular surface and an impingement lesion. This lesion initiated a fatigue crack that propagated to cause failure of the meniscal bearing. This is the first report of a meniscal bearing fracture without a posterior marker wire.