Simple and robust method for lithium traces determination in drinking water by atomic emission using low-power capacitively coupled plasma microtorch and microspectrometer
Related Experiment Video
Updated: May 8, 2026

Total Internal Reflection Absorption Spectroscopy (TIRAS) for the Detection of Solvated Electrons at a Plasma-liquid Interface
Published on: January 24, 2018
Andreea R Zsigmond1, Tiberiu Frentiu, Michaela Ponta
1Sapientia Hungarian University of Transylvania, Environmental Science Department, Matei Corvin 4, 400112 Cluj-Napoca, Romania.
Abstract:
A method for Li determination in drinking water using atomic emission spectrometry in a new low-power Ar capacitively coupled plasma microtorch (15 W, 0.6 L min(-1)) with a detection limit of 0.013 μg L(-1) was developed. The method is based on external calibration in the presence of a buffering solution containing 5 mg L(-1) Na, K, Ca, Mg added both to calibration standards and water samples. The statistical validation on 31 bottled drinking water samples (0.4-2140 μg L(-1) Li) using the Bland and Altman test and regression analysis has shown results similar to those obtained by the standard additions method. The buffering solution approach is simpler than the standard additions and has demonstrated good intra- and interday precision, accuracy and robustness. It was successfully applied over a wide concentration range of Li and multimineral matrix with a pooled precision of 2.5-3.5% and 99±9% accuracy.
Related Concept Videos
Atomic Emission Spectroscopy: Lab
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Atomic Emission Spectroscopy: Instrumentation
Gas Chromatography: Types of Detectors-II
Flame Photometry: Lab