Related Experiment Video
Updated: May 8, 2026

Use of a Percutaneous Ventricular Assist Device/Left Atrium to Femoral Artery Bypass System for Cardiogenic Shock
Published on: August 16, 2021
Direct thrombolytic therapy for thrombosis of a centrifugal flow left ventricular assist device
Ali Abdul Jabbar1, Raymond Yau, O H Frazier
1Department of Cardiology, Texas Heart Institute at St. Luke's Episcopal Hospital, Houston, Texas, USA.
Abstract:
In patients receiving left ventricular assistance for end-stage heart failure, pump-related thrombosis is a potentially fatal complication that may manifest as hemolysis and abnormally high pump power consumption. The optimal approach to treat this complication is unclear, especially in patients who are not heart transplant candidates or who have a high surgical risk. We report the successful use of direct thrombolytic therapy for recurrent hemolysis and thrombosis-related pump failure in a patient with the HeartWare ventricular assist device (HVAD), a third-generation, centrifugal flow pump. We know of only one previous case in which direct thrombolytic therapy has been used to treat the failure of a third-generation device. On the basis of our experience in the present case, we believe that direct thrombolytic therapy is a feasible option in these patients.
Related Concept Videos
Venous Thrombosis III: Interprofessional Care
Pulmonary Embolism II: Diagnostic Studies and Interprofessional Care
Cardiac Catheterization III: Left Heart Catheterization
Peripheral Artery Disease III: Interprofessional Care
Heart Failure VI: Adjunct Therapies